Emerson NGA 2000 TO2 Respiratory Product User Manual


 
Instruction Manual
760008-A
September 2001
Rosemount Analytical Inc. A Division of Emerson Process Management Operation 3-33
Model NGA 2000 TO2
a. Diagnostic Menus
This menu has two submenus for viewing
and resetting any software errors.
The “Control module diagnostics” menu is
not applicable to this analyzer model.
Analyzer Module Diagnostics
This menu displays any software error
messages of the analyzer.
Options:
Make a note of the error message.
Reset the message: Press the or key
in the “Edit to reset:” line. Choose “Reset”
with the or key and confirm with the .
The error messages will disappear if the
cause of the error no longer exists and
the tag “Report” will appear. If the error
message remains, contact customer
service.
Load/Save Module Configura-
tion
This menu provides several functions to
send or load configuration data of the
analyzer through the serial interface.
These functions are only available if an
SIO with serial interface is installed.
NOTE:
When loading configuration data all of
the current configuration in the mem-
ory will be overwritten.
This function allows the setting of flow
units for the current channel and provides
a display of the actual flow of the channel.
Send configuration to serial interface !
The configuration data in memory will be
sent through the serial interface of the
analyzer to an external computer or other
device.
Load configuration data from serial inter-
face !
Configuration data will be loaded into
memory from an external computer or
other device through the serial interface of
the analyzer. The current configuration in
memory will be overwritten.
Replace current configuration with factory
settings !
Deletes the configuration in memory and
re-establishes the factory default setting
from the Flash-EPROM.
TO2 37.50 ppm
Measure
Back…
-- Analyzer Module Diagnostics --
Last message: No error
And: No error
And: No error
And: No error
And: No error
And: No error
And: No error
And: No error
Edit to reset: Report
Software error code (1=no error): 1
TO2 37.50 ppm
Measure
Back…
-- Diagnostics Menus --
Control module diagnostics…
Analyzer module diagnostics…
<<< >>>
TO2 37.50 ppm
Measure
Back…
-- Load/Save Configuration (CM/MCA) --
Send configuration to seral interface!
Load configuration from serial interface!
- BE CAREFUL with this function -
Replace current configuration with factory settings!