Emerson NGA 2000 TO2 Respiratory Product User Manual


 
Instruction Manual
760008-A
September 2001
Rosemount Analytical Inc. A Division of Emerson Process Management Operation 3-19
Model NGA 2000 TO2
d. Linearization
Linearization is only performed when the
sensor is changed. See section 4-3 be-
low.
e. Measurement Display Configuration
Figure 3-20. Measurement Display Configuration
The “Signal on mini-bargraph” lines are
used to set various status values that can
display on the four lines available on the
single component (Measure) display. Se-
lect a line and press the key. To change
the value use the and keys. The val-
ues will scroll through the allowable se-
lections as follows:
Electrolyte level Displays “OK” or “Low.”
Range Displays the range number from 1 to 4.
Sample flow Displays the sample flow rate in ml/min with a bargraph.
Sensor current Displays the sensor current in uA with a bargraph.
Temp. current Displays the heater current in uA with a bargraph.
Sensor temp.
Displays the sensor temperature in °C with a bargraph.
Validity Not functional on the TO2.
Operation status Not functional on the TO2.
Move the cursor to the “Measurement display
configuration…” line and press the key.
The “Measurement Display Configuration”
menu will be displayed.
The first two lines allow for setting the con-
centration value displayed precision and
number of digits.
The “Module identification tag:” line set the
displayed neumonic for the analyzer on all
display screens. Select the line and press the
key to change the values using the and
keys to select a digit and then the and
keys to change the value. The values will
scroll through all numbers and upper and
lowercase letters.
TO2 37.50 ppm
-- Anal
y
zer Module Setu
p
and Controls --
Calibration…
Alarms…
Ranges…
Linearization…
Measurement display configuration…
Concentration measurement…
Flow measurement…
Temperature measurement…
Measure Back…
Status…
TO2 37.50 ppm
Measure
Back…
-- Measurement Display Configuration --
Displayed concentration digits: 6
Digits after decimal point: 2
Module identification tag: TO2
Signal on mini-bargraph – 1: Electrolyte level
Signal on mini-bargraph – 2: Temp current
Signal on mini-bargraph – 3: Range
Signal on mini-bargraph – 4: Sensor current