Emerson NGA 2000 TO2 Respiratory Product User Manual


 
Instruction Manual
760008-A
September 2001
3-16 Operation Rosemount Analytical Inc. A Division of Emerson Process Management
Model NGA 2000 TO2
Span Calibration
Depending upon the nature of your appli-
cation, it may be beneficial to verify the
span calibration of the analyzer module
every 3-4 months. The following proce-
dure illustrates how to initiate a span cali-
bration.
Figure 3-17. Span Calibration
Introduce a suitable span gas to the analyzer
and allow the reading to stabilize.
Move the cursor to the “Span gas:” line and
press the key. Enter the correct span gas
value by using the and keys to adjust the
value and the and keys to select the
digit. Press the key again to save the
value.
Move the cursor to the “Calibrate to span
gas!” line and press the key. The new span
calibration will be calculated.
TO2 37.50 ppm
-- Anal
y
zer Module Setu
p
and Controls --
Calibration…
Alarms…
Ranges…
Linearization…
Measurement display configuration…
Concentration measurement…
Flow measurement…
Temperature measurement…
Measure Back…
Status…
TO2 37.50 ppm
Measure
Back…
-- Calibration --
Gas scale factor…
Reset calibration!
Calibrate to span gas!
Span gas: 90.0 ppm
Status…
Main Menu
Analyzer and I/O, expert controls and setup…
Analyzer module controls… (Or Analyzer module setup…)