Emerson NGA 2000 TO2 Respiratory Product User Manual


 
Instruction Manual
760008-A
September 2001
Rosemount Analytical Inc. A Division of Emerson Process Management Operation 3-29
Model NGA 2000 TO2
Special Scaling for Concentra-
tion Signal
This menu allows for the setting of each
of the 8 outputs to be the same as the
range limit “Yes” or as set on the previous
menus.
See Section 3-5c, Ranges for setting the
range limits.
Pressing F5 (More…) changes to the
submenu Analog Output Updates per
Second.
Pressing the F5 (More…) key changes to
the submenu “Analog Output Updates per
Second.”
Analog Output Updates Per
Second
This menu allows for the setting of the
update rate for each of the 8 outputs.
c. Serial Interface Setup
From menu System SIO Module (Figure
3-25), select Serial interface setup…
In this menu, the parameters for data
transfer between the analyzer and exter-
nal devices are set. The choices in this
menu depend on the configuration of the
analyzer. The full specification of the se-
rial interface is described in its own man-
ual.
Options:
Baud rate: 300, 1200, 2400, 4800,
9600, 19200
Data bits: 7, 8
Stop bits: 1, 2
Parity: None, Even, Odd
Echo mode: Enabled, Disabled
Handshake: None, Xon/Xoff
Transmission delay: 0…100.
Type of installed serial inter-
face:
RS232
RS485/2w
RS485/4w
RS485/4w bus
None
TO2 37.50 ppm
Measure
Back…
-- Special Scaling for Concentration Signal --
(Scaling is the same as range limits)
Output #1: Yes
Output #2: Yes
Output #3: Yes
Output #4: Yes
Output #5: Yes
Output #6: Yes
Output #7: Yes
Output #8: Yes
More…
TO2 37.50 ppm
Measure
Back…
-- Serial Interface Setup --
Baud rate: 19200
Data bits: 8
Stop bits: 1
Parity: None
Echo mode: Disabled
Handshake: Xon/Xoff
Transmission delay: 0
Type of installed serial interface: RS232
Communication protocol: AK
Special protocol definitions…
TO2 37.50 ppm
Measure
Back…
-- Analog Output Updates per Second --
Output #1: 10
Output #2: 0
Output #3: 0
Output #4: 0
Output #5: 0
Output #6: 0
Output #7: 0
Output #8: 0