Emerson NGA 2000 TO2 Respiratory Product User Manual


 
Instruction Manual
760008-A
September 2001
3-28 Operation Rosemount Analytical Inc. A Division of Emerson Process Management
Model NGA 2000 TO2
Output(s) value on analyzer
failure
Choose the desired signal level to cause
a failure condition. The choices are:
Actual
BeginOfRange
EndOfRange
BeginOfRange-10%
BeginOfRange+10%
Output number
Choose the output number (1-8) for set-
ting the fine adjustment.
Operation mode
Normal: The absolute measurement sig-
nal will be sent to the analog output.
Adjust 0V: Used to set the display equal
to the analog output for 0V and 0 mA. Life
zero signals (4-20 mA and 2-10V) are set
automatically and cannot be adjusted.
Adjust 10V: Used to set the display equal
to the analog output for 10V and 20 mA.
Select the “Fine adjustment for 0% out-
put” and/or “Fine adjustment for 100%
output” lines with the or key. Adjust
to the desired value with the or key
and confirm with the key. The range of
values are:
3000 to 6000 for 0% (default 4096)
600 to 1000 for 100% (default 819)
The last three lines of the “Analog Output
Setup” menu are display only for configu-
ration values of the analog output.
Signal name: The name of the sig-
nal chosen in the “Choose signal”
menu.
Current signal value: The current
value of the variable.
Source module: The name of the
module chosen in the “Choose sig-
nal source module” menu.Pressing
the F5 (More…) key changes to the
submenu “Special Scaling for Con-
centration Signal.”
This menu allows for the setting of each
of the 8 outputs to be the same as the
range limits “Yes” or as set on the previ-
ous menus.
See section 3-5c for setting the range lim-
its.
Pressing F5 (More…) changes to the
submenu “Special Scaling for Concentra-
tion Signal”.
TO2 37.50 ppm
Measure Back…
-- Output Signal if Assigned Module Fails --
Output(s) value on analyzer failure: BeginOfRange – 10%
-- Fine Adjustment –
Output number: 1
Operation mode: Normal
Fine adjustment for 0% output: 4096
Fine adjustment for 100% output: 819
More…
TO2 37.50 ppm
Measure Back…
-- Output Signal if Assigned Module Fails --
Output(s) value on analyzer failure: BeginOfRange – 10%
-- Fine Adjustment –
Output number: 1
Operation mode: Normal
Fine adjustment for 0% output: 4096
Fine adjustment for 100% output: 819
More…
TO2 37.50 ppm
Measure Back…
-- Output Signal if Assigned Module Fails --
Output(s) value on analyzer failure: BeginOfRange – 10%
-- Fine Adjustment –
Output number: 1
Operation mode: Normal
Fine adjustment for 0% output: 4096
Fine adjustment for 100% output: 819
More…