OxyBalance Oxygen Display
and Averaging System
Current Selecting the Current tab on the right side of the touch screen will display the
Current screen. The two tabs at the top of the screen will display the Probes
and Averages information.
Probes
The Probes tab, shown in Figure 3-6, displays the status and the oxygen
percentage for each of the eight possible probes. The color of the bar denotes
the status of the probe signal.
Green = Probe is OK
Blue = Probe is in calibration
Red = Probe has failed
Yellow = Analog input signal is outside the normal range, but not far
enough to be considered failed.
NOTE
If a probe signal level is not shown in green, then it is not being used in any
average calculations.
Figure 3-6 provides an example of a Probes tab with probes in several of
these conditions. Probe 1 is in Fault condintion, and probe 5 is in calibration.
Neither of these probes will be used in any average calculations.
Figure 3-6. Probes Tab