Agilent Technologies 6843A Pacemaker User Manual


 
Performance Tests
15
Table 2-4. (Continued) Performance Test Record – Agilent 6834B AC Power Source
Model: Agilent 6843B Report No.: Date:
Test Description Min Spec Results Max Spec Meas
Uncert **
Voltage Programming & Readback Accuracy
High Range@ 45 Hertz
Program 300V
Front Panel Display
299.250
Vrms-400mV
______V
______V
300.750
Vrms+400mV
140mV
( 1.2mV )
High Range@ 400 Hertz
Program 300V
Front Panel Display
298.200
Vrms-400mV
______V
______V
301.800
Vrms+400mV
140mV
( 1.2mV )
High Range@ 1000 Hertz
Program 300V
Front Panel Display
296.700
Vrms-400mV
______V
______V
303.300
Vrms+400mV
140mV
( 1.2mV )
Low Range@ 45 Hertz
Program 150V
Front Panel Display
149.475
Vrms-325mV
______V
______V
150.525
Vrms+325mV
70mv
( 0.6mV )
Low Range@ 400 Hertz
Program 150V
Front Panel Display
148.950
Vrms-325mV
______V
______V
151.050
Vrms+325mV
70mV
( 0.6mV )
Low Range@ 1000 Hertz
Program 150V
Front Panel Display
148.200
Vrms-325mV
______V
______V
151.800
Vrms+325mV
70mV
( 0.6mV )
CV Load Effect Vo-0.75V ______V Vo+0.75V
CV Line Effect Vo-0.3V ______V Vo+0.3V
rms Current Readback Io-0.030A ______A Io+0.030A l.lmA
Total Harmonic Distortion
Audio Analyzer
Front Panel Display
0%
0%
______%
______%
1%
1%
Frequency Programming & Readback Accuracy
Program 45Hz 44.985 ______Hz 45.014 0.9mHz
Front Panel Display Fo-0.014Hz ______Hz Fo+0.014Hz
Program 1 KHz 999.89 ______Hz 1000.110 0.010Hz
Front Panel Display Fo-0.110Hz ______Hz Fo+0.110Hz
**Note: 1. Measurement uncertainties are only valid when using test equipment listed in Table 2-1.
2. Voltage Programming and Readback measurements uncertainties are for a 3458A DMM in the 1000 volt range.
3. Measurement uncertainties in parenthesis are only if a 30:1 ratio transformer is used with the 3458A DMM