Teledyne 100EU Blood Glucose Meter User Manual


 
Model 100EU Instruction Manual Addendum Appendix C – Warranty Repair Questionnaire
05930 Rev A C-1
Appendix C – Warranty Repair Questionnaire
CUSTOMER: _________________________________ PHONE: __________________________________________________
CONTACT NAME: ____________________________ FAX NO. __________________________________________________
SITE ADDRESS: _________________________________________________________________________________________
MODEL 100EU SERIAL NO.: ___________________ FIRMWARE REVISION: _________________
____________________
1. A
RE THERE ANY FAILURE MESSAGES? _______________________________________________________________
________________________________________________________________________________________________________
________________________________________________________________________________________________________
2. PLEASE C
OMPLETE THE FOLLOWING TABLE: (NOTE: DEPENDING ON OPTIONS INSTALLED, NOT ALL TEST
PARAMETERS BELOW WILL BE AVAILABLE IN YOUR INSTRUMENT)
Parameter
*IF OPTION IS INSTALLED
Recorded Value Acceptable Value
RANGE
PPB/PPM
STABIL
PPB
0.3 PPB WITH ZERO AIR
STABIL2
PPB <= 1 PPB WITH ZERO AIR
PRESS
IN-HG-A AMBIENT (- 2) IN-HG-A
SAMPLE FLOW
cm
3
/MIN 650 ± 10%
PMT SIGNAL WITH
ZERO AIR
m
V -20 TO 150 mV
PMT SIGNAL AT
SPAN GAS CONC
mV
PPB/PPM
0-5000 mV
0-20000 PPB
NORM PMT AT SPAN
GAS CONC
mV
PPB/PPM
0-5000 mV
0-20000 PPB
UV LAMP
mV 2000 TO 4000 mV
UV STAB
mV < 1
5mV
LAMP RATIO
mV 30 T
O 120%
STR. LGT
PPB 10
0 PPB/ ZERO AIR
DARK PMT
mV 200-325
DARK LAMP
mV -50 TO 200 mV
SLOPE
1.0 ± 0.5
OFFSET
mV < 250 mV
HVPS
V 40
0 – 800
RCELL TEMP
ºC 50ºC ± 1
BOX TEMP
ºC AMBIENT ± 5
PMT TEMP
ºC 9.5ºC ± 2º CONSTANT
IZS TEMP*
ºC 50ºC ± 1
ETEST
mV 2000 mV ± 500
OTEST
mV 2000 mV ± 1000
Values are in the Signal I/O
REF_4096_MV
mV 4096mv±2mv and Must be Stable
REF_GND
mV 0± 0.5 and Must be Stable