Model 100EU Instruction Manual APPENDIX A - Version Specific Software Documentation
A-22 05928 Rev F.0B
PRINTED DOCUMENTS ARE UNCONTROLLED DCN 5063
Table A-3: M100EU Test Functions, Revision F.0B
Name
1
Message Text Description
RANGE RANGE= 500.0 PPB
3
SO2 RNG=500.0 PPB
3, 10, 11
D/A range in single or auto-range modes.
RANGE1 RANGE1=50 0.0 PPB
3
SO2 RN1=500.0 PPB
3, 10, 11
D/A #1 range in independent range mode.
RANGE2 RANGE2=50 0.0 PPB
3
SO2 RN2=500.0 PPB
3, 10, 11
D/A #2 range in independent range mode.
CO2RANGE CO2 RNG=100 PCT
11
CO
2
range.
O2RANGE O2 RNG=100 WT%
10
O
2
range.
STABILITY ST ABIL=0.0 PPB
3
SO2 STB=0.0 PPB
3, 10
O2 STB=0.0 WT%
10
CO2 STB=0.0 PCT
11
Concentration stability #1. Configure with
SETUP-MORE-STBL.
STABILITY2
6
ST ABIL2=0.0 PPB
3
SO2 STB2=0.0 PPB
3, 10
O2 STB2=0.0 WT%
10
CO2 STB2=0.0 PCT
11
Concentration stability #2. Configure with
SETUP-MORE-STBL.
RESPONSE
2
RSP= 1.11(0.00) SEC
Instrument response. Length of each
signal processing loop. Time in
parenthesis is standard deviation.
VACUUM
5
VAC=9.1 IN-HG-A Vacuum pressure.
SAMPPRESS PRES=29.9 IN-HG-A Sample pressure.
SAMPFLOW SAMP FL=700 CC/M Sample flow rate.
PMTDET PMT=762.5 MV Raw PMT reading.
NORMPMTDET NORM PMT=742.9 MV
PMT reading normalized for temperature,
pressure, auto-zero offset, but not range.
UVDET UV LAMP=3457.6 MV UV lamp reading.
STABILITYUV
6
UV STB=5.607 MV UV lamp stability reading.
LAMPRATIO LAMP RATIO=100.0 %
UV lamp ratio of current reading divided
by calibrated reading.
STRAYLIGHT STR. LGT=0.1 PPB Stray light offset.
DARKPMT DRK PMT=19.6 MV PMT dark offset.
DARKLAMP DRK LMP=42.4 MV UV lamp dark offset.
SLOPE SLOPE=1.061
Slope for current range, computed during
zero/span calibration.
OFFSET OF FSET=250.0 MV
Offset for current range, computed during
zero/span calibration.
CO2SLOPE
11
CO2 SLOPE=1.0000
CO
2
slope, computed during zero/span
calibration.
CO2OFFSET
11
CO2 OFFSET=0.00 %
CO
2
offset, computed during zero/span
calibration.
O2SLOPE
10
O2 SLOPE=0.980
O
2
slope, computed during zero/span
calibration.
O2OFFSET
10
O2 OFFSET=1.79 %
O
2
offset, computed during zero/span