Teledyne 100EU Blood Glucose Meter User Manual


 
TELEDYNE INSTRUMENTS
Model 100EU Instruction Manual
APPENDIX A - Version Specific Software Documentation
05928 Rev F.0B A-27
DCN 5063 PRINTED DOCUMENTS ARE UNCONTROLLED
AGND_DARK
3
7 AGND reading during dark cycles
AGND_LIGHT
3
8 AGND reading during light cycles
VREF_DARK
3
9 VREF4096 reading during dark cycles
VREF_LIGHT
3
10 VREF4096 reading during light cycles
Rear board primary MUX analog inputs
PMT detector
HVPS_VOLTAGE 1 HV power supply output
PMT_TEMP 2 PMT temperature
UVLAMP_SIGNAL 3 UV lamp intensity
4 Temperature MUX
PHOTO_ABS
8
5 Pre-amplified UV lamp intensity
O2_SENSOR
10
6 O
2
concentration sensor
SAMPLE_PRESSURE 7 Sample pressure
TEST_INPUT_8 8 Diagnostic test input
REF_4096_MV 9 4.096V reference from MAX6241
SAMPLE_FLOW 10 Sample flow rate
VACUUM_PRESSURE
2
10 Vacuum pressure
CO2_SENSOR
11
11 CO
2
concentration sensor
12–13 Spare (thermocouple input?)
14 DAC MUX
REF_GND 15 Ground reference
Rear board temperature MUX analog inputs
BOX_TEMP 0 Internal box temperature
RCELL_TEMP 1 Reaction cell temperature
IZS_TEMP 2 IZS temperature
3 Spare
O2_CELL_TEMP
10
4 O
2
sensor cell temperature
TEMP_INPUT_5 5 Diagnostic temperature input
TEMP_INPUT_6 6 Diagnostic temperature input
7 Spare
Rear board DAC MUX analog inputs
DAC_CHAN_1 0 DAC channel 0 loopback
DAC_CHAN_2 1 DAC channel 1 loopback
DAC_CHAN_3 2 DAC channel 2 loopback
DAC_CHAN_4 3 DAC channel 3 loopback
Rear board analog outputs
CONC_OUT_1,
DATA_OUT_1
0 Concentration output #1 (SO
2
, range #1),
Data output #1
CONC_OUT_2,
DATA_OUT_2
1 Concentration output #2 (SO
2
, range #2),
Data output #2
CONC_OUT_3
10,
DATA_OUT_3
2 Concentration output #3 (CO
2
or O
2
),
Data output #3