Leica DM4000B Microscope & Magnifier User Manual


 
43
6. Startup
• Focus the direct image with the collector
(46.6).
Use the adjusting buttons on the rear side of
the lamp housing (46.2,46.4) to pivot the arc’s
mirror image to the side or completely out of
the beam path. The lamp filament’s focused
image remains visible (Fig. 50).
Use the adjusting buttons (46.1) and (46.5) to
place the direct arc image right or left on an
imaginary center line of the centering plane
(Fig. 51).
Then pivot the arc’s mirror image with the ad-
justing knobs (46.2 and 4) and focus it using
the reflector (46.3).
Use the adjusting knobs (46.2 and 4) to orient
the mirror image symmetrically to the direct
image (Fig. 52).
Defocus the image with the collector knob
(46.6) until the arc image and mirror image are
no longer recognizable and the image is
uniformly illuminated.
Exchange the reflector cube for lamp
adjustment for the original filter cube.
Centering the Hg 100 W and Xe 75 W
mercury lamps
In the adjustment window, you see the direct
arc image and the mirror image, which in most
cases are shifted together.
Fig. 52 Direct arc image and mirror image in target
position (in reality, the image is less focused)
Fig. 51 Direct arc image in target position
(in reality, the image is less focused)
Fig. 50 Direct arc image focused but decentered
(in reality, the image is less focused)