Leica DM4000B Microscope & Magnifier User Manual


 
21
5. Assembly
5.6 Light Sources for the Incident Light Axis
Caution:
During assembly, always unplug the power
supply unit of the 106 z lamp housing from its
socket.
Never touch the glass parts of the burner
with bare hands.
Never look directly into the beam path (blind-
ing hazard).
During assembly work on xenon burners, al-
ways wear the supplied protective gloves and
face protection (Fig. 15) (risk of explosion).
106 z lamp housing
This lamp housing is used with a 12V 100W halo-
gen lamp or various gas discharge lamps.
Inserting the 12V 100W halogen lamp into the
106 z lamp housing
Unscrew the fastening screws of the cover
and lift up the cover (16.1).
Unscrew the fastening screws of the lamp
mount (16.8) and pull out the mount (Fig. 17).
Fig. 15
Protective gloves and mask
Fig. 16 106 z lamp housing (on the side, open)
1 Cover raised
2 Collector
3 12 V 100 W lamp or
gas discharge lamp in mount
4 Reflector (mirror)
5, 6, 7 Adjusting screw for x-y reflector
8 Fastening screw for lamp mount
9 Socket for contact plug
1
2
3
898
4
5
6
7