Reference Manual
00809-0100-4027, Rev AA
December 2005
Rosemount 4500
A-4
Alarm Configuration
Rosemount
High Alarm: ≥ 21.75 mA
Low Alarm: ≤ 3.75 mA
Custom Level
(1)
High Alarm: 20.2 - 23.0 mA
Low Alarm: 3.6 - 3.8 mA
Humidity Limits
0-100% relative humidity
Physical Specifications Process Connections
•1
1
/2 inch Tri-Clamp Connection
• 2 inch Tri-Clamp Connection
•1
1
/2 inch Fractional Line Connection
Process-Wetted Parts
Process Isolating Diaphragms
316L SST
Hastelloy C-276
®
Surface Finish
15 μ-inch Ra mechanically polished
10 μ-inch Ra electropolished
Non-Wetted Parts
Electronics Housing
304 SST
NEMA 4X
IP 66
Surface Finish
32 μ-inch Ra mechanically polished
Sensor Module Fill Fluid
Neobee M-20
Shipping Weight for Rosemount 4500
3.0 lb. (1.36 kg.)
(1) Low alarm must be 0.1 mA less than low saturation and high alarm must be 0.1 mA greater