Emerson 4500 Pacemaker User Manual


 
Reference Manual
00809-0100-4027, Rev AA
December 2005
2-3
Rosemount 4500
Figure 2-2. HART Installation
Flowchart
START HERE
Bench
Calibration?
Field Install
No
Configure
(Section 2)
Set Units
Set Range
Points
Set Output
Type
Set Damping
Verify
Apply Pressure
Yes
Within
Specifications
?
Yes
No
Refer to
Section 4
Troubleshooting
Check Switches
(page 2-5)
Mount
Transmitter
(pages 2-4)
Wire Transmitter
(pages 2-7–2-8)
Power
Transmitter
(page 2-7)
Trim Transmitter
for Mounting
Effects
(page 4-5)
Done
Confirm
Transmitter
Configuration
(page 3-18)
Confirm
Transmitter
Configuration
(page 3-18)