Digi DC-190 Scale User Manual


 
Scale Programming 31
7.2 Item Code Maintenance
Item code maintenance includes viewing item codes in memory and deleting item memory.
View Item Codes in Memory
The following procedure allows viewing all of the item codes stored in memory but will not change any of the
information (unit weight, tare weight, part number, quantity, part name, setpoints, and lot number) stored with
these codes.
1. Press the
MODE key. The WEIGHT display shows ProG and the QUANTITY display shows C XX.
2. Press the
CODE key. The QUANTITY display shows the first item code programmed and stored in
memory, example:
Id 123.
3. Press the
+ key to view the next item in memory and continue pressing the + key until the DC-190 beeps.
SPEC 7,
bits 0,1
Setpoint Type Setpoint Designation
00 Percent quantity Setpoint 1: Quantity
The quantity value can be no greater than 999999.
Setpoint 2: Percent of Setpoint 1
The percentage value can be no greater than 999%. The value for Setpoint 2 is calculated
by multiplying the value for Setpoint 1 by the percentage value entered for Setpoint 2.
NOTES:
If Setpoint 1 is equal to 999999, then Setpoint 2 must be set less than or equal to
100%.
Setpoints 3–6 are not available when Setpoint 2 is a percent setpoint.
01 Percent weight Setpoint 1: Weight
The quantity value can be no greater than 999999.
Setpoint 2: Percent of Setpoint 1
The percentage value can be no greater than 999%. The value for Setpoint 2 is calculated
by multiplying the value for Setpoint 1 by the percentage value entered for Setpoint 2.
NOTE:
If Setpoint 1 is equal to 999999, then Setpoint 2 must be set less than or equal to
100%.
Setpoints 3–6 are not available when Setpoint 2 is a percent setpoint.
10 Upper and lower
quantity limit
Setpoint 1: Quantity
Setpoint 2: Quantity
Setpoints 3–6: Quantity
11 Upper and lower
weight limit
Setpoint 1: Weight
Must be a weight value less than or equal to the capacity of the scale.
Setpoint 2: Weight
Must be a weight value less than or equal to the capacity of the scale, but must be less
than Setpoint 1 value.
Setpoints 3–6: Weight
Must each specify a weight value less than the value of the preceding setpoint.
NOTES:
All weight-based setpoint values must be entered with a decimal and all trailing zeros. All percentage-based
setpoint values must be entered to the nearest whole number (no fractional percentages).
The DC-190 can program up to six setpoints by repeating Steps 2 through 6, SPEC 18, bit 0 through 2 (Number of
Setpoints) determines the number of setpoints. The six setpoints are TTL output for quantity or weight, but not
percent quantity or percent weight. These values may be programmed 1 through 6 low to high or 1 through 6 high to
low.
Table 7-1. Setpoint Programming for Quantity and Weight