4.4 RS-232 Specifications (143 Settings)
To configure RS-232 specifications, press and hold the REZERO key and enter 143 using the numeric keypad.
SPC40 appears in WEIGHT display and configuration of four bits (for SPEC 40) appears in QUANTITY display.
Follow the procedure described in Section 4.1 on page 17 to update the SPEC settings for the 143 access code.
SPECs 40–42 and SPEC 43, bits 0–2, represent the selectable RS-232 output fields.
SPECs 46–59 are not used. All bits for these SPECs must be set to 0.
SPEC Bit 3 Bit 2 Bit 1 Bit 0
40 Lot Number
0: Yes
1: No
Memo (Part Name)
0: Yes
1: No
Alternative Part Number
0: Yes
1: No
ID Code
0: Yes
1: No
41 Net Weight
0: Yes
1: No
Gross Weight
0: Yes
1: No
Setpoint
0: Yes
1: No
Inventory
0: Yes
1: No
42 Date and Time
0: Yes
1: No
Quantity & Total Quantity
0: Yes
1: No
Unit Weight
0: Yes
1: No
Tare Weight
0: Yes
1: No
43 Autoprint within Setpoints
0: No
1: Yes
Non-stable Ouput
0: No
1: Yes (allow serial output
with scale in motion)
Status Output
0: Yes
1: No
Scale Number
0: Yes
1: No
44 Clear Operator Name
Each Use (BCP-300)
0: No (operator name held
in register for multiple
inputs)
1: Yes
Company Name
(BCP-300)
0: Default only
1: Input with bar board
Manual Print with Setpoints
0: Yes
1: No
Autoprint Overrange
Setpoint
0: No
1: Yes
45 Not used Not used Print with Zero Quantity
0: No
1: Yes
Batch Print Once
0: No
1: Yes
46–59 Not used: set to 0 Not used: set to 0 Not used: set to 0 Not used: set to 0
Table 4-3. DC-190 143 Settings