(C) Example: Programming the Monitor for Four (4) Periods
Suppose that the start time of the monitoring session is 09:00 and you would like the monitor to
measure the patient's blood pressure every 30 minutes until 22:00, and from 22:00 to 23:00 you
want the monitor OFF so that the patient may sleep, and from 23:00 to 06:00 you would like the
patient's blood pressure measured once an hour, and finally from 06:00 you want the blood pressure
measured every 15 minutes. Remember that the Time for Period One (1) also defines the End of the
monitoring session. The end time for Period Four (4) will default to the Time for Period One (1), in
this case 09:00.
PERIOD TIME INTERVAL DURATION
1 09:00-22:00 30 minutes 13 hours
2 22:00-23:00 OFF 1 hours
3 23:00-06:00 1 hour 7 hours
4 06:00-09:00 15 minutes 3 hours
Total: 24 hours
The following is the recorder setup according to these instructions.
Note: Set Periods One (1), Two (2), Three (3) and Four (4). Leave the other Periods as
“-”.
PERIOD START INTERVAL
1 9 30 minutes
222OFF
3 23 1 hour
4 6 15 minutes
5--
6--
Figure 25: Monitor programming example 3