Monthly Maintenance 1 2 3 4 5 6 7 8 9 10 11 12
Exchange the cleaning solutions
Replace 7.3 buffer installed for
30 days
Replace Cal G/L reagent installed for
30 days
Inspect the capillary seal
Bimonthly Maintenance 1 2 3 4 5 6
Replace 6.8 buffer and Wash/Zero
reagents installed for 60 days
Replace sample tubing
Quarterly Maintenance 1234
Clean the sample chamber
Clean the hemolyzer
RAPIDLAB 800 MAINTENANCE SCHEDULE Month/Year ________________