System Administration 5Ć47
6. Press Yes.
The QC File Setup screen appears with all data fields empty, as shown in
Figure 5-28.
        
   
    
7. Complete the form.
         
the keypad type the low and high target limits for each parameter. Press
Enter after you complete each field.
the optional bar
code scanner
scan the target range bar code for each parameter from QC
Expected Values. If you scan the wrong bar code, a message
appears prompting you to scan the correct parameter bar code.
Press
OK and continue.
When you complete both fields for a parameter, the system automatically
calculates the target value and the action range.
8. If a second screen of parameters exists, press
Next Screen and enter values on
that screen.