3Ć32 Rapidlab 800 Operator's Manual
 
Perform the following procedures every 6 months. The semiannual maintenance
described in this section is based on analyzing 30 samples per day. If your
laboratory analyzes more than 30 samples per day, perform this maintenance more
frequently.
   
Materials required:
S air filter
1. Push the air filter cover down until it snaps out and then pull the cover away
from the system as shown in Figure 3-17.
   
 
  
2. Remove the old air filter and discard it.
3. Press the new air filter into place.