Use this procedure to define the number of significant digits reported for certain
primary parameters. Table 5-13 lists the parameters that have high and low
resolution options. The default value is high.
  
    
Low pH 0.01
H
+
nmol/L 1
pCO
2
mmHg
kPa
1
0.1
pO
2
mmHg
kPa
1
0.1
High pH 0.001
H
+
nmol/L 0.1
pCO
2
mmHg
kPa
0.1
0.01
pO
2
mmHg
kPa
0.1
0.01
1. Access the System Options screen from the Menu screen:
a. Select 6 System Setup and press Enter.
b. Select
5 System Options and press Enter.
The System Options screen appears with the cursor in the Reporting
Resolution frame, as shown in Figure 5-24.
