The monthly maintenance schedule is based on the expiration date of the opened
reagents.
Effective cleaning is accomplished by alternating Cleaning Solution 1 (C1) and
Cleaning Solution 2 (C2) each month.
C1 is on your system replace it with C2.
C2 is on your system replace it with C1.
WARNING
Wear safety glasses, gloves, and a laboratory coat when handling the
reagents.
1. At the Ready screen, press
Enter.
The Reagent Levels screen appears.
2. Remove the cleaning solution bottle from the reagent manifold as shown in
Figure 3-7.
3. Write the date installed in the space provided on the new cleaning solution
bottle.