Tektronix TDS 520A Blood Glucose Meter User Manual


 
Measurement System
TDS 620A, 640A, & 644A User Manual
Gated Measurements
The gating feature lets you limit measurements to a specified portion of the
waveform. When gating is Off, the oscilloscope makes measurements over
the entire waveform record.
When gating is activated, vertical cursors are displayed. Use these cursors to
define the section of the waveform you want the oscilloscope to measure.
This is called the
gated region
.
1. Press MEASURE Gating (main) Gate with V Bar Cursors (side)
(see Figure 3-47).
Figure 3-47: Measure Menu — Gating
2. Using the general purpose knob, move the selected (the active) cursor.
Press SELECT to change which cursor is active.
Displaying the cursor menu and turning V Bar cursors off will
not
turn
gating off. (Gating arrows remain on screen to indicate the area over
which the measurement is gated.) You must turn gating off in the Gating
side menu.