Leica RM2245 Microscope & Magnifier User Manual


 
60
Instruction manual V 1.4 – 02/2007
7. Customized solutions
Feed motion no longer occurs.
Problem Corrective actionPossible cause
1. STEP mode is still active (LED in
SECT button lights up).
1. The end of the residual path has
been reached.
2. The specimen was already
within the residual path when
the instrument was switched on.
The specimen in the residual
path area cannot be returned
using the coarse feed buttons.
1. Press the TRIM/SECT button to
switch to trimming mode. Return
the specimen using the coarse
feed button.
1. Select the trim mode by pressing
TRIM/SECT. Then return the
specimen using the coarse feed
button. Then retract the specimen
using the coarse feed button.
2. Follow the procedure in Item 1.
No display; no reaction to
buttons pressed after the
instrument is switched on.
1. The power cable is not connect-
ed correctly.
2. Power fuses are defective.
3. The cable of the control panel is
not connected correctly.
4. The voltage selection switch is
set incorrectly.
1. Check the connection of the
power cable.
2. Replace power fuses (see 8.3).
3. Check the connection of the
cable to the control panel.
4. Check the voltage setting and
correct it if necessary
(see Chapter 4.4).
7.1.2 Malfunctions, possible causes and troubleshooting
Following is a list of the most common problems which can arise while working with the
instrument, along with possible causes and troubleshooting procedures.
If the malfunction cannot be remedied with any of the options listed in the table, or the problem
occurs repeatedly, inform the Leica Service Support immediately.