Leica RM2245 Microscope & Magnifier User Manual


 
17
Leica RM2245
Setting up the instrument
Do not transport the instrument by holding it on the handwheel
shaft, the object head or the cover!
By holding the instrument (5) at the front by the baseplate, and at the
rear by the recessed grip (Fig. 8), lift it out of the molded cushion (6) of
the packaging and place it on a stable laboratory table.
4.2 Installation site requirements
Stable, vibration-free laboratory bench with horizontal and even stage
plate; practically vibration-free floor.
No other instruments nearby which might cause vibrations.
Room temperature permanently between +10°C and +35 °C.
Obstruction-free access to the handwheel.
Never operate the instrument in rooms with explosion hazard.
Fig. 8
9
6
5
The two sliding elements (9) located on the
rear of the base plate make it easier to move
the instrument on the table.
To move the instrument, hold it by the front
of the base plate, lift it up gently and slide it
on its slides.
4. Startup