Leica RM2245 Microscope & Magnifier User Manual


 
30
Instruction manual V 1.4 – 02/2007
Setting up the knife holder base
Release the clamping lever (50) by rotating it
counterclockwise.
Insert the knife holder base (51) using the
notch (52) on the bottom into the T-piece (55)
of the microtome base plate (53).
To secure the knife holder base, turn the
clamping lever (50) clockwise.
The knife holder base (51) can be moved back
and forth on the microtome base plate. This
allows bringing the knife holder to optimal
sectioning position in relation to the specimen.
There is a scale (54) on the right side of the
microtome base plate. This enables faster and
better positioning of the knife holder at the
specimen if various combinations of standard
specimens and specimen holders are used. The
rear edge of the knife holder base (51) functions
as the scale reference.
Fig. 20
55
50
51
52
53
5.2 Inserting the knife holder
Inserting the knife holder
Loosen the screw (58) using an Allen key
size 4 (71) until the knife holder (57) can be
moved.
Place the knife holder (57) with the under-
side groove onto the T-piece (56) of the knife
holder base (51).
To clamp, retighten the screw (58).
Fig. 21
Enlarged detail:
Scale for better
repositioning of the
knife holder for
varying specimen
heights.
54
5. Operation
56
57
58
71