Leica RM2245 Microscope & Magnifier User Manual


 
26
Instruction manual V 1.4 – 02/2007
Configuring the retraction settings
To call up the retraction settings, press the MENU MODE and CLEAR
buttons simultaneously.
The current set value is displayed as a three-digit number in the four-
digit display (e.g. "025" = 25 μm).
Select the desired retraction value.
The retraction value can be adjusted in increments of 5 μm to a maxi-
mum of 100 μm using the
- buttons on the control panel.
To exit the retraction settings, press MENU MODE.
A retraction movement takes place by the newly selected value after
each section.
While the specimen is in retraction, the yellow LED on the RETRACT
display (4) lights up.
Switching off the specimen retraction
To call up the retraction settings, press MENU MODE and CLEAR
simultaneously.
To switch off retraction, press the
button on the control panel until
the display indicates "OFF".
To exit the retraction settings, press MENU MODE.
When retraction is switched off, the specimen is not retracted.
The yellow LED (4) of the RETRACT indicator does not light up.
5. Operation
+
or
4
+
4