Leica RM2245 Microscope & Magnifier User Manual


 
59
Leica RM2245
7. Customized solutions
Certain instrument malfunctions appear on the 3-digit LED display
or are indicated through illuminated LED’s on the control panel.
7.1.1 Error reports
Indication E1
If this message follows the software version indication after switching on
the instrument, a loss of all parameters has occurred.
Press any key.
All parameters are then set back to the default values.
Now all settings must be checked to see if the previously entered
values are still correct.
The values may have to be reentered.
You can continue with your normal working routine.
If E-1 appears repeatedly when switching the instrument on, the lithium
battery may need replacement (service life approx. 7 years).
Proceed as described above.
Work normally after checking the settings.
Have the battery replaced by Leica service as soon as possible.
LEDs (40), (41) in the coarse feed buttons are lit at the same time.
This indicates a defect in the detection of the STOP and HOME positions
of the object head.
The problem can only be resolved by customer service.
7.1 Instrument malfunctions
40
41