SECTION 12—BATTERIES
Part No 1149267 65 TDX® SC TDX® Spree
Replacing the Batteries
NOTE:RefertoDisconnectingBatteryCablesonpage 65.
Disconnecting/Connecting Battery Cables
Disconnecting Battery Cables
ƽ WARNING
The use of rubber gloves is recommended when working with batteries.
NEVER allow any of your tools and/or battery cables to contact both battery
terminals at the same time. An electrical short may occur and serious personal
injury or damage may occur.
NOTE:Forthisprocedure,refertoFIGURE 12.6onpage 67.
1. Removethebatteries.RefertoRemoving/InstallingtheBatteries
onpage 57.
2. Cutthetiewrapthatsecuresthebatteryterminalcapinplace(Detail“B”of
FIGURE 12.6).
3. Slideterminalcapsupontothebatterycables(FIGURE 12.6).
4. DisconnectPOSITIVE(+)batterycablefromthePOSITIVE(+)batteryterminal
(FIGURE 12.6).
5. DisconnectNEGATIVE()batterycablefromNEGATIVE()batteryterminal
(FIGURE 12.6).