Fluke 35040 Light Therapy Device User Manual


 
Appendix
Operation Instructions
B
B-13
Cause: The input current has exceeded the 1.200 uA limit.
Action: Stop the exposure. Change the setup to reduce the current.
Explanation: The instrument has a 1.000 uA limit on measurable current and will continue to measure
charge for currents up to 1.200 uA. The instrument can accept an input current up to 3 mA without
damage to the instrument.
B.7 Customization
The screens shown in this section represent the appearance of the Customization software running within
a MS-DOS® window running on Windows® 95/98.
B.7.1 The Main Menu
The Main menu appears after customization information has been received from the Model 35040
Therapy Dosimeter. The menu items are Ion Chambers, Special Items, Measurement, Bias Settings,
Print, and Exit.
B.7.2 The Measurement Menu
The Measurement menu provides items Displays, Timer Intervals, and Quit. Select Displays to view the
Measurement Displays, Timer Intervals to view the Measurement Timer Intervals, or Quit to return to the
Main menu.
Measurement Displays
The Model 35040 Therapy Dosimeter can contain up to sixteen (16) measurement displays, which are
viewed from the Measurement screen. From the Measurement Displays section of the Customization
software, you may customize the appearance of any of the measurement displays. All measurement
displays must show the annunciators on the first line. You may choose what appears on the other three
lines
Press RESET/MEASURE
Reduce exposure and
ADC
Current Overrange