Emerson 3812 Oxygen Equipment User Manual


 
94 Using a Field Communicator to configure the meter
Section 4: Configuration Daniel 3812 Liquid Ultrasonic Flow Meter Installation Manual
May 2015 3-9000-765 Rev H
Procedure
1. Remove electrical power to the meter. If installed, remove the endcap security latches
and seals and then, remove the endcap.
2. Refer to the Field Communicator Users Manual wiring diagrams and commissioning
instr
uctions provided with your handheld device. Register the product to activate the
end user license.
3. Fully charge the Field Communicator battery prior to use. Imp
o
rtant: Do not change
the battery in a hazardous area environment. The power supply is not intrinsically safe.
4. On the meter, run the wires through the field wiring conduit and into the transmitter
ele
ctronics enclosure.
Figure 4-19 3812 transmitter field wiring conduit entries
A. Field wiring conduit entries (4)
A.