Bosch Appliances F.01U.252.722 Home Care Product User Manual


 
24 en | Programming NurseCall Relay Unit
F.01U.252.722 | V1 | 2011.12 User Manual Bosch Security Systems
c. Choose 0 to 9 characters for the criterion, followed by 0 to 9 spaces.
Default values are 2 characters + 1 space.
7. Repeat the operation for
a. the floor (0 to 3 characters and 0 to 9 spaces), by default 2 characters and 1 space
b. the room (0 to 3 characters and 0 to 9 spaces), by default 2 characters and 1 space
c. the bed (0 or 1 character), by default 1 character.
4.4.5 Selection "DISPLAY NORSK"
if you have selected DISPLAY NORSK, press to confirm.
1. Set the rotational display of alarms.
2. Select YES or NO with and . Default value is YES. Confirm with .
3. Configure the display of the 6 identification positions. The outcome is visible at the top.
Select a field with and , change a value with and .
–Confirm with .
–Choose 0 to 9 characters for the criterion, followed by 0 to 9 spaces.
Default values are 2 characters and 1 space.
4. Repeat the operation for
the floor (0 to 3 characters), by default 2 characters
the room (0 to 3 characters), by default 3 characters
the bed (0 or 1 character), by default 1 character.
NOTICE!
Configuration of the display is possible only in display mode FL,RO,BE (floor, room, bed).
If the NurseCall system is in display mode NUMBER, there is no need of further programming.
NOTICE!
The display mode NUMBER ZZZZ (alphanumeric display) is not compatible with the Display
Norsk. Make sure to choose another display mode when connecting a Display Norsk to the
Relay Plus Unit. See the NurseCall Main Unit user manual for more information.
¬ RS-232 Output
DISPLAY NORSK ¢
¬ DISPLAY NORSK
Rota.of Alarms:YES ¡
DISPLAY : ffrrrb
FL: 2 +RO: 3 +BE:1
NOTICE!
Configuration of the display is possible only in display mode FL,RO,BE (floor, room, bed).
If the NurseCall system is in display mode NUMBER, there is no need of further programming.