Bayer HealthCare Blood Glucose Monitoring System Blood Glucose Meter User Manual


 
35
Alternate
95% Confidence Interval
Site Opr. n y = Slope / Intercept r
Palm Lay 40 1.00x – 0.06 0.95 to 1.07 –0.63 to 0.24 0.986
Forearm Lay 40 0.96x – 0.04 0.89 to 1.04 –0.73 to 0.72 0.973
Abdomen Lay 39 0.85x + 0.63 0.76 to 0.95 –0.09 to 1.14 0.958
Thigh Lay 40 0.97x – 0.23 0.88 to 1.07 –0.95 to 0.34 0.962
Opr. = Operator; Lay = Lay User
C. Results from the Abdomen D. Results from the Thigh
0
5
10
15
20
25
30
051015 20 25 30
Ascensia C
ONTOUR
Fingerstick Result (mmol/L)
Ascensia C
ONTOUR
Abdomen Result (mmol/L)
0
5
10
15
20
25
30
151050202530
Ascensia C
ONTOUR
Fingerstick Result (mmol/L)
Ascensia C
ONTOUR
Thigh Result (mmol/L)
y = x
y = x
Passing and Bablok Regression Statistics / Pearson correlation