Multiple Site Testing Evaluation
Forty persons with diabetes, in steady state for glucose,
performed self-capillary blood glucose assays using the
Ascensia™ C
ONTOUR™ System and one reagent lot. A
fingerstick was performed with the Ascensia™ M
ICROLET
®
Adjustable Lancing Device. Samples from the palm, forearm,
thigh, and abdomen were then obtained, and those results
were compared to the Ascensia C
ONTOUR results obtained
from the finger. One subject was unable to obtain a sample
from the abdomen. The blood glucose range from the
fingerstick measurements was 2.7 to 22.6 mmol/L, with an
average of 9.2 mmol/L. The hematocrit range was 36 to 52%,
with an average of 42%.
34
A. Results from the Palm
0
5
10
15
20
25
30
05101520 25 30
Ascensia C
ONTOUR
Fingerstick Result (mmol/L)
Ascensia C
ONTOUR
Palm Result (mmol/L)
B. Results from the Forearm
0
5
10
15
20
25
30
151050202530
Ascensia C
ONTOUR
Fingerstick Result (mmol/L)
Ascensia C
ONTOUR
Forearm Result (mmol/L)
y = x
y = x