Bayer HealthCare Blood Glucose Monitoring System Blood Glucose Meter User Manual


 
27
B. Health Care Professional Results
0
5
10
15
20
25
30
05101520 25 30
Plasma Comparative Method Glucose Result (mmol/L)
Ascensia CONTOUR Result (mmol/L)
Lot 22B Lot 22C Lot 23B
Reagent
95% Confidence Interval
Lot Opr. n y = Slope / Intercept r
22B Lay 50 1.03x – 0.00 0.93 to 1.15 –1.17 to 0.85 0.947
HCP 50 1.01x – 0.28 0.91 to 1.11 –1.22 to 0.61 0.967
22C Lay 51 1.07x – 0.14 1.00 to 1.16 –0.79 to 0.45 0.969
HCP 51 1.04x – 0.08 0.99 to 1.10 –0.47 to 0.39 0.972
23B Lay 101 1.00x + 0.14 0.95 to 1.04 –0.23 to 0.54 0.973
HCP 101 0.98x + 0.20 0.94 to 1.05 –0.27 to 0.62 0.960
Opr. = Operator; Lay = Lay User; HCP = Healthcare Professional
Passing and Bablok Regression Statistics / Pearson correlation