Bayer HealthCare Blood Glucose Monitoring System Blood Glucose Meter User Manual


 
Precision
1. Laboratory Precision: To assess the repeatability (within-
run precision) of the Ascensia
CONTOUR
Blood Glucose
Monitoring System, 100 blood glucose readings (10
readings on each of 10 instruments) were obtained with
venous blood specimens at five glucose concentrations. The
following table summarizes the mean and pooled within-
Meter %CV at each level.
2. Control Testing Precision: Precision results were obtained
from duplicate assays of Control Solutions at two diabetes
clinics with the Ascensia C
ONTOUR System. Testing was
performed over 63 days. People with diabetes and
healthcare professionals (HCP) at each clinic evaluated two
reagent lots. One lot was evaluated at both clinics. The
tables below show the combined results for the lay users
and HCPs. The overall standard deviations (SD) and
coefficients of variation (CV) include the within run and
between run / day variations.
23
Duplicate Overall Overall
Reagent Assays Mean SD CV
Lot Operator (n = ) (mmol/L) (mmol/L) (%)
22B HCP (n = 1) 35 2.34 0.074 3.2
Lay User (n = 51) 76 2.35 0.102 4.3
22C HCP (n = 4) 39 2.39 0.063 2.6
Lay User (n = 53) 78 2.39 0.093 3.9
23B HCP (n = 5) 74 2.36 0.063 2.7
Lay User (n = 104) 154 2.32 0.112 4.8
Performance Evaluations
(for Healthcare Professionals)
Low Control Solution
Mean 2.6 mmol/L 5.1 mmol/L 8.0 mmol/L 13.1 mmol/L 20.9 mmol/L
%CV 2.4% 2.5% 2.8% 3.1% 2.4%