21
What You See What it Means What You Should Do
HI Test result is above Wash your hands and test site. Repeat
33.3 mmol/L the test. If the test result is still
(600 mg/dL) “HI,” contact your physician or
healthcare professional (HCP)
immediately.
CAUTION: Glucose levels above
13.9 mmol/L (250 mg/dL) may
indicate a potentially serious medical
condition.
Control Test • Control Solution is Check all expiration dates. Do not
result is out-of- past expiration date use expired testing materials. Run
range (too high or is past the 3-month another Control Test. If still out-of-
or too low). open use date. range, retest with a new Test Strip
• Test Strip is past and Control Solution. If still out of
expiration date or is range, call the Bayer Diabetes
past the 6 month Helpline 1-800-268-7200.
open use date.
• Deteriorated Test
Strip due to heat,
cold or exposure to
moisture.
• Control Solution not
at room temperature.
Some display Possible meter elec- Replace batteries. If display is
segments do tronics or battery fail- still incorrect, call the Bayer
not appear ure. Compare display Diabetes Helpline 1-800-268-7200.
during self-test. with picture on inside This may affect the way you
of Front Cover. see your results.