Bayer HealthCare Blood Glucose Monitoring System Blood Glucose Meter User Manual


 
fingerstick. Following the Ascensia™ CONTOUR™ System
assays, a blood sample was collected for comparative
laboratory glucose and hematocrit determinations. The
glucose concentration of the samples from the two sites
ranged from 1.7 to 23.3 mmol/L with an average of 10.1
mmol/L. The range of hematocrits was 28 to 55% with an
average of 44%.
26
A. Lay User Results
0
5
10
15
20
25
30
05
1
0
15 20 25 30
Plasma Comparative Method Glucose Result (mmol/L)
Ascensia C
ONTOUR
Result (mmol/L)
Lot 22B Lot 22C Lot 23B
y=x