Programming Commands
Common Commands
*RST
The *RST (reset) command returns the Agilent 86120C to a known
condition.
Syntax
*RST
Description For a listing of reset conditions, refer to the following table. This com-
mand cannot be issued as a query. Since this command places the
instrument in single measurement acquisition mode, any current data
is marked as invalid and a measurement query such as :FETCh?
results in error number –230, “Data corrupt or stale”. You must ini-
tiate a new sweep with :INIT:IMM before you can use the :FETCh com-
mand.
Table 4-15. Conditions Set by *RST Reset (1 of 2)
Item Setting
Display mode single wavelength
Start wavelength 1270 nm
Stop wavelength 1650 nm
Graphical display off
Measurement acquisition single
Wavelength calibration vacuum
Elevation correction value 0 meters
Wavelength units nm
Amplitude units dBm
Power offset 0 dB
Peak threshold 10 dB
Peak excursion 15 dB
Measurement speed normal