Philips M3860A Blood Pressure Monitor User Manual


 
Philips Medical Systems
B
B-1
B Technical Specifications
The specifications for the HeartStart FR2+ provided in this chapter apply to
both the M3860A and M3861A, unless otherwise noted. Additional
information can be found in the
Technical Reference Manual for the FR2
Defibrillator, located online at www.medical.philips.com.
HeartStart FR2+ Defibrillator Specifications
Physical
Environmental
category nominal specifications
Size 2.6" high x 8.6" wide x 8.6" deep (6.6 cm x 21.8 cm x 21.8 cm).
Weight Approximately 4.7 lbs (2.1 kg) with M3863A FR2 standard battery installed.
Approximately 4.5 lbs (2 kg) with optional M3848A FR2+ rechargeable battery
installed.
category nominal specifications
Operating Temperature
and Humidity
32° to 122° F (0° to 50° C).
0% to 95% relative humidity (non-condensing).
Standby Temperature
and Humidity
32° to 109° F (0° to 43° C).
0% to 75% relative humidity (non-condensing).
Applies to HeartStart FR2+ with battery installed and stored with defibrillator
pads.
Altitude Meets MIL-810E 500.3, Procedure II (-500 feet to 15,000 feet).
Shock/Drop Abuse
Tolerance
Meets MIL-STD-810E 516.4, Procedure IV (after a 1 meter drop to any edge,
corner, or surface, in standby mode).
Vibration Meets MIL-STD-810E 514.4-17.
Sealing With data card tray and battery installed, meets IEC 529 class IP54.
ESD Meets EN 61000-4-2:1998 Severity Level 4.
EMI (Radiated) Meets EN 60601-1-2 limits (1993), method EN 55011:1998 Group 1 Level B.