Setup and Advanced Mode Features
6-5
Philips Medical Systems
6
Using Setup Features
NOTE: To move around the menus displayed, use the Option buttons as
follows:
• Press the
LOWER Option button to move the highlight bar from one
item to another on the menu.
• Press the
UPPER Option button to select the highlighted item or to
scroll through the settings for that item.
The FR2+ comes with a factory default setup designed to meet the needs of
most users. The setup feature of the FR2+ lets you review the current setup
of your HeartStart FR2+ or install a revised setup if appropriate. To go to the
SETUP menu:
1. Remove and reinstall the battery to bring up the first menu on the screen.
NOTE: This screen will not be displayed if the FR2+ is connected to
defibrillator pads (that are applied to the patient) when the battery is
inserted, and you will not be able to access the menu items. In addition,
the battery insertion selftest and periodic automatic selftests cannot run
while the defibrillator pads are connected. Be sure to unplug the pads
connector from the FR2+ after each use. Do not store the FR2+ with
the pads connected.
resume key
ON, OFF OFF
Enables (ON) or disables (OFF) user-initiated interruption
of the CPR pause and return to analyzing. If either the
CPR Timer or the NSA Action setting is programmed to
1.5 minutes or longer, the Resume Key setting is
automatically enabled (
ON).
If enabled, the Resume
Key is accessed by
pressing the lower Option
button indicated by an
arrow on the FR2+
display, as shown in the
sample screen:
advanced use
prompt
interval
(minutes)
0.5, 1.0, 1.5,
2.0, 2.5, 3.0
0.5 Sets the interval for “Press to Analyze” prompts provided
during advanced mode operation.
parameter settings default description