Philips M3860A Blood Pressure Monitor User Manual


 
Maintaining, Testing, and Troubleshooting Your HeartStart FR2+
4-13
Philips Medical Systems
4
Voice prompts:
ANALYZING INTERRUPTED
or
CANNOT ANALYZE
or
STOP ALL MOTION
The patient is being moved or jostled.
Radio or electrical sources are
interfering with ECG analysis.
The environment is dry and movement
around the patient is causing static
electricity to interfere with ECG
analysis.
Stop CPR; do not touch the
patient. Minimize patient motion. If
the patient is being transported,
stop the vehicle if needed.
Check for possible causes of
radio and electrical interference
and remove them from the area.
Responders and bystanders
should minimize motion,
particularly in dry environments
that can generate static electricity.
Voice and screen prompts:
NO SHOCK DELIVERED
The patient impedance is not appropriate
for the FR2+ to deliver a shock.
Make sure the defibrillator pads
are correctly positioned on the
patient according to the diagram
on the back of the pads.
Make sure the defibrillator pads
connector is completely inserted
in the connector socket.
Press the defibrillator pads firmly
to the patient's chest.
Replace the defibrillator pads if
necessary.
Voice prompt:
SHOCK BUTTON NOT
PRESSED
Shock has been advised but not delivered
within 30 seconds. (FR2+ has been
disarmed.)
When next prompted, press the
Shock button to deliver shock.
symptom possible cause recommended action