Philips M3860A Blood Pressure Monitor User Manual


 
M3860A/M3861A HEARTSTART FR2+
7-2
Philips Medical Systems
To install a data card:
1. Make sure the data card is clean and dry.
2. Load the data card into its plastic tray, with the tray’s “tongue” fitting
over the matching yellow area on the data card. The label on the card
should face up. The label has an arrow indicating which side to insert
into the data card port.
3. Make sure the FR2+ is off (in standby mode), or that the battery has
been removed.
4. Hold the loaded tray by its handle and gently insert the tray into the data
card port on the right side of the FR2+. Push the tray all the way into the
port until only the tab remains outside the FR2+ case. Do not force the
tray into the port. If the tray is hard to insert, remove it and make sure that
the arrow label is face up and pointing toward the data card port.
The data card will automatically record incident data the next time the
HeartStart FR2+ is turned on.
To avoid running out of data card space during an incident, it is
recommended that each data card be used to record the information for only
one incident and that it be replaced after each use of the FR2+.
If you record information from more than one incident on a data card, it is
important to review how much time is left on the used data card before
recording a new incident. To do this, load the data card into the data card
tray, insert the tray in the FR2+, then remove and reinstall the battery. The
first screen displayed shows how much recording time remains on the card.
NOTE: During an incident, if for any reason you turn off the FR2+ for less
than five minutes, the FR2+ considers this to be a “continued use” situation,
and:
the information stored about the incident is saved,
additional events recorded after the device is turned back on will be
treated as part of the same incident, and
the selftest will not automatically run if the battery is replaced.
IMPORTANT NOTE: Do not remove the battery while incident data are
being recorded to a data card. To ensure that no incident data are lost, turn
the FR2+ off (return it to standby mode) before replacing the battery.