Philips M3860A Blood Pressure Monitor User Manual


 
M3860A/M3861A HEARTSTART FR2+
4-10
Philips Medical Systems
TESTS — how many tests have been run. Four figures are shown: daily
(upper left), weekly (upper right), and monthly (lower left) periodic
selftests, and battery insertion selftests (lower right).
REV — device language, model, and software revision.
Battery History
Information about use of the battery currently installed in your FR2+ is also
available. To review the history of the battery, select
NEXT from the menu
screen displayed when you insert the battery, then select
BATTERY HISTORY
from the next menu displayed.
The battery history information is read from the internal memory of the battery.
It includes:
USE MINUTES — the total operating time (in minutes), including selftest
time, for this battery;
CHARGES — the total number of full defibrillation charges that have been
provided by this battery, including selftest charges;
BATTERY — a GOOD BATTERY (M3863A) or a fuel gauge display
(M3848A) showing 25%, 50%, 75% or 100%, or a
LOW BATTERY or
REPLACE BATTERY message, as appropriate.
STATUS — the current status of this battery, displayed in a binary code.
Make a note of this code if technical service is needed.