Philips M3860A Blood Pressure Monitor User Manual


 
Maintaining, Testing, and Troubleshooting Your HeartStart FR2+
4-15
Philips Medical Systems
4
Status Indicator:
SOLID RED X
Audio signal:
NONE
The battery is missing or completely
depleted.
The training & administration pack is
being used in the administration
function (the solid red X is normal in
this case) or has been left in the FR2+
by mistake.
A self-test detected a failure.
Install a new M3863A FR2
standard battery or a charged
M3848A FR2+ rechargeable
battery in the FR2+ and perform
the battery insertion test (BIT).
Remove the training &
administration pack and install a
battery.
Remove and reinstall the battery
and perform the battery insertion
self-test. If it fails, install a new
M3863A FR2 standard battery or
a charged M3848A FR2+
rechargeable battery and repeat
the test. If it fails again, do not use
the FR2+.
Status Indicator:
SOLID RED X
Audio signal:
CHIRPING
The training & administration
pack is being used in the
ADMINISTRATION function
and more than 10 minutes have
passed without user interaction
(button press or pads change).
The training & administration pack is
being used in the
TRAINING function
and more than 30 minutes have
passed without user interaction
(button press or pads change).
To continue using the training &
administration pack, press any
button (except On/Off).
To return the FR2+ to standby
mode, remove the Pack and install
a battery.
Status Indicator:
NONE
The FR2+ has been physically damaged. Check for visible damage. Do not
use the FR2+ if it appears to be
damaged.
Remove and reinstall the battery to
perform the battery insertion
self-test. If it fails, install a new
M3863A FR2 standard battery or
a charged M3848A FR2+
rechargeable battery and repeat
the test. If it fails again, do not use
the FR2+.
symptom possible cause recommended action