Linear PERS-4200 Series Medical Alarms User Manual


 
35
17. Appendix A - Reporting Codes
EVENT
DESCRIPTION
REPORTING CODE
CID
4 X 2 SIA
Group Partition (Ext)
CONSOLE EMERGENCY ALARM 1100 00 69 QA
ALARM CANCEL (ABORT) 1406 00 70 AB
CONSOLE LOW BATTERY 1302 00 73 YT
CONSOLE BATTERY RESTORE 3302 00 71 YR
CONSOLE SUPERVISORY (RECEIVER BOARD TROUBLE) 1381 00 40 MS
CONSOLE SUPERVISORY RESTORE 3381 00 40 MJ
LOSS OF AC POWER 1301 00 72 AT
AC POWER RESTORE 3301 00 94 AR
HOME MODE (ACTIVITY MONITOR ON)
[1]
1372 00 91 UR
AWAY MODE (ACTIVITY MONITOR OFF)
[1]
1371 00 92 UB
INACTIVITY ALARM (NO ACTIVITY) 1102 00 90 NA
AUTOMATIC STATUS (AUTOMATIC TEST) 602 00 93 RP
AUTOMATIC STATUS (TEST OFF NORMAL) 608 00 95 RY
REMINDER MESSAGE TIME OUT
[1]
1356 00 96 MI
HIGH TEMP ALARM (HEAT) 1158 00 74 KA
HIGH TEMP RESTORE 3158 00 75 KR
LOW TEMP ALARM (FREEZE) 1159 00 76 ZA
LOW TEMP RESTORE 3159 00 77 ZR
FIRE (SMOKE) ALARM 1111 XX
[2]
10 + SENSOR # FA
FIRE (SMOKE) LOW BATTERY
[1]
1384 XX
[2]
40 + SENSOR # FT
FIRE (SMOKE) SUPERVISORY 1200 XX
[2]
40 + SENSOR # FS
FIRE (SMOKE) SUPERVISORY RESTORE 3200 XX
[2]
40 + SENSOR # FQ
CO ALARM (GAS) 1162 XX
[2]
10 + SENSOR # GA
CO LOW BATTERY 1384 XX
[2]
40 + SENSOR # GT
CO SUPERVISORY 1381 XX
[2]
40 + SENSOR # GS
CO SUPERVISORY RESTORE 3200 XX
[2]
40 + SENSOR # GQ
EMERGENCY ALARM (PERSONAL/MEDICAL) 1101 XX
[2]
10 + SENSOR # MA
EMERGENCY LOW BATTERY 1384 XX
[2]
40 + SENSOR # MT
EMERGENCY SUPERVISORY 1381 XX
[2]
40 + SENSOR # MS
EMERGENCY SUPERVISORY RESTORE 3381 XX
[2]
40 + SENSOR # MJ
TWO WAY AUDIO REQUEST 606 00 N/A
[3]
L90
Reporting Code Notes:
• 1
Optional report depending on programming.
• 2
The CID Group/Partition # (Ext) code is the zone number that the alarm
or event is reported against. Console and non-zone-specifi c alarms/events
are reported using CID group/partition #00
• 3
The 4x2 format does not have a two-way audio request code. The call
must be established manually.