3.1.1 Technical Data for the VT1200 S
General data:
Cutting frequency (± 10 %): ......................................................................................................................... 85 Hz (± 10%)
Amplitude: ............................................................................................................ from 0 – 3mm, in 0.05mm increments
Cutting speed (± 10 %): ..............................................................................................................................0.01 – 1.5mm/s
Return speed (± 10 %): .........................................................................................1.0 – 5mm/s, in 0.5mm/s increments
Total vertical specimen stroke: ......................................................................................................... 20 mm (motorized)
Cutting range: ............................................................................................................................................................. 45 mm
Cutting window: ......................................................................................................................................... 0.5mm – 45mm
Specimen retraction: ............................................................................... 0 – 100μm (adjustable, can be deactivated)
Maximum specimen size:
With standard blade holder: ...................................................................................................................... 33 x 50 mm
Specimen orientation, rotating: ........................................................................................................................... 360°
Specimen plate, swiveling:...............................................................................................................................0 – 10°
Section thickness adjustment: ........................................... manual in 1μm increments or automatic max. 1000μm
Ambient conditions:
Operating temperature range: .................................................................................................... min. 10°C – max. 35°C
Relative humidity:................................................................................................................................................. max. 60%
Storage temperature: ............................................................................................................................................ 5 – 55°C
Storage humidity: ....................................................................................................................................................... < 60%
Height: ........................................................................................................................................................ up to 2000 m NN
Electrical data:
Rated voltage range (± 10 %): ....................................................................................................................... 100V – 240V
Rated frequency (± 10 %): .................................................................................................................................... 50/60 Hz
Power consumption:................................................................................................................................................... 35 VA
Protection class: .................................................................................................................................................................. I
Power fuses: .................................................................................................................................................... T 1A L 250 V
Pollution degree: ................................................................................................................................................................. 2
Overvoltage category: ........................................................................................................................................................ II
Electrical overload protection: ..................................................................................................................................... yes
Internal current limit for the electronics: .................................................................................................................... yes
Dimensions:
3. Instrument Characteristics
L x W x H: ................................................................................................................................ 600mm x 250mm x 230mm
Height with magnifier support: ............................................................................................ 600mm x 250mm x 320mm
Height with microscope: ....................................................................................................... 600mm x 250mm x 490mm
Control unit (when the bases are folded in): ....................................................................... 190mm x 150mm x 72mm
Weight:
Without magnifier support and control unit: ........................................................................................................... 56 kg
VT1200S control unit ...................................................................................................................................................... 1 kg
Magnifier support: ......................................................................................................................................................... 2 kg
Microscope support with stereomicroscope:........................................................................................................... 4 kg