Leica Leica VT1200S Microscope & Magnifier User Manual


 
13
Leica VT1200 and VT1200S – Microtome
Sets the blade feed rate.
Select the values using
setting dial 1.
Sets the amplitude.
Select the values
using setting dial 1.
Moves the speci-
men into the lowest
position, (LED is
illuminated when
the lowest position
is reached)
Starts or stops the sectioning
process. In semiautomatic
mode, the sectioning process
stops immediately; in automatic
mode, the sectioning process
is finished completely.
Setting dial 1, for blade
feed rate (SPEED),
AMPLitude or section
thickness for automatic
mode (AUTOFEED)
Setting dial 2, for
section thickness and
moving the specimens
vertically
Selects the step size
in 1μm,10μm or
100μm
increments
Fig. 11
Sets the section thickness
addition display μm to "0".
Interrupts the sectioning
process in automatic mode.
Pressing the button again
reactivates the sectioning
process.
ON/OFF lamp
Possible in automatic
sectioning mode only.
Select the desired section
thickness for the automatic
feed using setting dial 1.
Toggles in and out of the
menu. Memory for 8 para-
meter sets.
Toggling between single
stroke (SINGLE) and continu-
ous stroke (CONT) in automatic
mode is possible; in semiauto-
matic mode, only single stroke
(SINGLE) is possible.
1st cutting window edge,
LED illuminates when the
window edge is set.
2nd cutting window edge,
LED illuminates when the
window edge is set.
3. Instrument Characteristics
3.3.1 VT1200S Control Panel
Toggles between automatic (AUTO) and
semiautomatic (MAN) sectioning mode
Moves the specimen
upwards (as long as the
button is pressed). When
the uppermost position is
reached, the LED lights up.
Moves the
blade
towards the
specimen
Moves the
blade away
from the
specimen
Caution: Press the button
until the desired position is
reached. When the end
position is reached, the LED
lights up.
The Leica VT1200S is a fully automatic microtome with a vibrating blade. It can be operated in automatic
as well as semiautomatic sectioning mode.
In semiautomatic sectioning mode, a manual feed to the desired section thickness must be carried out be-
fore each cut. There is no automatic specimen retraction in this mode; however, retraction can be per-
formed manually.
In automatic mode, an automatic feed (AUTO FEED) to the selected section thickness is carried out before
each cut, and the specimen is lowered to the desired retraction value after each cut to prevent the speci-
men surface and the blade from coming into contact while the blade is being retracted.