7. Malfunctions: Meanings and Troubleshooting
Error/INF No. Error Troubleshooting Remark
- "Bottom" limit switch
of the Z-axis is not
reached during initial-
ization or operation.
- The "top" limit switch
of the Z-axis is not
reached.
- Both X sensors activat-
ed (during initialization
or in normal operation)
- Both Z sensors activat-
ed (during initialization
or in normal operation)
Error 28
- Check whether an obsta-
cle is blocking the travel
path of the ice bath.
Remove the obstacle.
If there is no obstacle,
and the error message
persists after the in-
strument is switched on
again: inform service.
- Check whether an obsta-
cle is blocking the travel
path of the ice bath.
Remove the obstacle.
If there is no obstacle,
and the error message
persists after the in-
strument is switched on
again: inform service.
- Inform service!
- Inform service!
Error 31
Error 32
After all error messages, the instrument MUST be switched off, then on
again using the main switch.
Error 27