Leica Leica VT1200S Microscope & Magnifier User Manual


 
44
Operating Manual V1.0 – 10/2006
Error/INF No. Error Troubleshooting Remark
7. Malfunctions: Meanings and Troubleshooting
- Basic calibration of
light barrier is not pos-
sible for VibroCheck.
- VibroCheck blade
search not successful.
- Communication error
between control unit
and VT.
- The buffer tray was not
pushed in up to the
rear pin or was not
clamped down during
installation.
- The VibroCheck was
not pushed all the way
in or was not clamped
down during installa-
tion.
- Switching from SINGL
to CONT is not possible.
- The cutting window
cannot be defined.
- Emitter diode or receiver
is defective.
- VibroCheck is defective.
- Inform service!
- No blade and/or blade
holder swung out or not
installed.
- Emitter diode or receiver
dirty. Clean
- VibroCheck is defective.
- VT1200S: display lights up,
but remains blank.
- VT1200: row of dots
flashes.
- Inform service!
Push in the buffer tray as
far as it will go and clamp
it down with the lever.
- Push in the VibroCheck as
far as it will go and clamp
it down with the lever.
Error 52
Error 53
- VT1200 drives to the
lowest Z position us-
ing the stepper motor
after an unsuccessful
search. (Allows blade
installation)
Communication
between keyboard
– VT1200/S
- The knife/blade
collides with the
buffer tray
- The VibroCheck
collides with the
blade
- In MAN mode, only single cuts are possible.
- In MAN mode, only single cuts are possible.
- If multiple cuts are required, switch to AUTO mode.